Details
The Fluke 718Ex offers: ATEX II 1G EEx ia IIC T4 Compliant.
Built-in pressure / vacuum hand pump, with fine adjust vernier and bleed valve.
30, 100, and 300 psi ranges (2, 7, and 20 bar)
Pressure measurement to 0.05% of full span, using an internal pressure sensor.
Pressure measurement to 200 bar using any of eight intrinsically safe Fluke 700PEx Pressure Modules.
Wide range of selectable measurement units for pressure.
Current measurement with 0.02% accuracy and 0.001 mA resolution.
Pressure switch test function.
Min-Max Hold functions.
Compact size and weight.
Simple, push-button user interface.
Rugged and reliable, ideal for field use.
Measurement Accuracy
Pressure Fluke 718Ex 30G
Range: -830 mbar to 2 bar (internal sensor)
Over pressure: 3x FS
Resolution: 0.0001 bar
Accuracy: 0.05% FS
Functions: Zero, Min, Max, Hold, Damp
Note: for use with non-corrosive gasses
Fluke 718Ex 100G
Range: -830 mbar to 7 bar (internal sensor)
Over pressure: 2x FS
Resolution: 0.001 bar
Accuracy: 0.05% FS
Functions: Zero, Min, Max, Hold, Damp
Note: for use with non-corrosive gasses
Fluke 718Ex 300G
Range: -830 mbar to 20 bar (internal sensor)
Over pressure: 375 psi, 25 bar
Resolution: 0.0001 bar
Accuracy: 0.05% FS
Functions: Zero, Min, Max, Hold, Damp
Note: for use with non-corrosive gasses
Fluke 718Ex series
via pressure module connector
Range: 29 pressure modules, 2.5 mbar to 700 bar
Over pressure per pressure module specs
Resolution: Per pressure module specs
Accuracy: To 0.025% of full span, per pressure module specs
Functions: Zero, Min, Max, Hold, Damp
Note: media compatibility per pressure module specs
Source Pressure, built in pump
Range: -820 mbar to full scale
Note: Supported Pressure Units PSI, in. H2O (4 °C), in. H2O (20 °C), kPa, cm H2O (4 °C),
cm H2O (20 °C), bar, mbar, kg/cm², mmHg, in Hg
Technical Data
Measure Accuracy: Range: 0-24 mA
Resolution: 0.001 mA
Accuracy: 0.02% rdg ± 2 counts
Display LCD, 5 digit pressure and current simultaneous
Environmental Specifications
Operating Temperature -10ºC to 55ºC
Storage Temperature -40ºC to 60ºC
Humidity (Without Condensation) 95% (10ºC to 30ºC)
75% (30ºC to 40ºC)
45% (40ºC to 50ºC)
35% (50ºC to 55ºC)
Operating Altitude 3,000 m max
Safety Specifications
Safety ATEX II 1G EEx ia IIC T4 compliant
CSA Certification I.S. Class I,
Div 1 Groups A-D , T4
Mechanical & General Specifications
Size 216 mm L x 94 mm W x 66 mm H (with holster)
Weight Approx. 992 g (with holster)
Warranty Three years, one year for pump
Shock 1 meter drop test
Vibration Random, 2 G, 5-500 Hz
Battery type One 9-volt alkaline battery ANSI/NEDA 1604A or IEC 6LR619V
Battery life: 4 to 20 hours, depending on functions used
Built-in pressure / vacuum hand pump, with fine adjust vernier and bleed valve.
30, 100, and 300 psi ranges (2, 7, and 20 bar)
Pressure measurement to 0.05% of full span, using an internal pressure sensor.
Pressure measurement to 200 bar using any of eight intrinsically safe Fluke 700PEx Pressure Modules.
Wide range of selectable measurement units for pressure.
Current measurement with 0.02% accuracy and 0.001 mA resolution.
Pressure switch test function.
Min-Max Hold functions.
Compact size and weight.
Simple, push-button user interface.
Rugged and reliable, ideal for field use.
Measurement Accuracy
Pressure Fluke 718Ex 30G
Range: -830 mbar to 2 bar (internal sensor)
Over pressure: 3x FS
Resolution: 0.0001 bar
Accuracy: 0.05% FS
Functions: Zero, Min, Max, Hold, Damp
Note: for use with non-corrosive gasses
Fluke 718Ex 100G
Range: -830 mbar to 7 bar (internal sensor)
Over pressure: 2x FS
Resolution: 0.001 bar
Accuracy: 0.05% FS
Functions: Zero, Min, Max, Hold, Damp
Note: for use with non-corrosive gasses
Fluke 718Ex 300G
Range: -830 mbar to 20 bar (internal sensor)
Over pressure: 375 psi, 25 bar
Resolution: 0.0001 bar
Accuracy: 0.05% FS
Functions: Zero, Min, Max, Hold, Damp
Note: for use with non-corrosive gasses
Fluke 718Ex series
via pressure module connector
Range: 29 pressure modules, 2.5 mbar to 700 bar
Over pressure per pressure module specs
Resolution: Per pressure module specs
Accuracy: To 0.025% of full span, per pressure module specs
Functions: Zero, Min, Max, Hold, Damp
Note: media compatibility per pressure module specs
Source Pressure, built in pump
Range: -820 mbar to full scale
Note: Supported Pressure Units PSI, in. H2O (4 °C), in. H2O (20 °C), kPa, cm H2O (4 °C),
cm H2O (20 °C), bar, mbar, kg/cm², mmHg, in Hg
Technical Data
Measure Accuracy: Range: 0-24 mA
Resolution: 0.001 mA
Accuracy: 0.02% rdg ± 2 counts
Display LCD, 5 digit pressure and current simultaneous
Environmental Specifications
Operating Temperature -10ºC to 55ºC
Storage Temperature -40ºC to 60ºC
Humidity (Without Condensation) 95% (10ºC to 30ºC)
75% (30ºC to 40ºC)
45% (40ºC to 50ºC)
35% (50ºC to 55ºC)
Operating Altitude 3,000 m max
Safety Specifications
Safety ATEX II 1G EEx ia IIC T4 compliant
CSA Certification I.S. Class I,
Div 1 Groups A-D , T4
Mechanical & General Specifications
Size 216 mm L x 94 mm W x 66 mm H (with holster)
Weight Approx. 992 g (with holster)
Warranty Three years, one year for pump
Shock 1 meter drop test
Vibration Random, 2 G, 5-500 Hz
Battery type One 9-volt alkaline battery ANSI/NEDA 1604A or IEC 6LR619V
Battery life: 4 to 20 hours, depending on functions used